Introduction
MAP-201 is specially designed for the food and pharmaceutical industries and can simultaneously detect oxygen in modified atmosphere packaging. It can be applied to all food industries and some pharmaceutical industries. This product is very suitable for the rapid and accurate detection of modified atmosphere packaging products in packaging lines, supermarkets and laboratories.
MAP-201 can measure the content of O2 very quickly in a few seconds. It is quite simple to use, without adjusting the flow rate, without temperature compensation, without considering cross-sensitivity and other technical factors.
Characteristic
Portable headspace analyzer
Rapid and Accurate Detection of Oxygen/Carbon Dioxide
Full Chinese interface, intuitive user interface, no training and direct use
Simple calibration steps
Large touch screen for easy reading
High Precision Full Range Detection Range
10 sets of data storage
Inspection needle placement hole, easy to carry
Typical applications
Fresh meat
Seafood
Cooked Food
Dairy
Fast Food
Bake
Salad
Coffee
Pharmacy
Glove box
agricultural research
Controlled storage
welding
Other
Working principle
oxygen
The core component of MAP-201 is an electrochemical oxygen sensor. The principle of the sensor is similar to that of a battery. When gas passes through the sensor, the sensor can generate current corresponding to the gas concentration. The current value can be converted to the concentration value and displayed on the display screen. The sensor has excellent strength and stability, high repeatability and fast response speed. Even if the oxygen concentration is high, the sensor will not be saturated, so the application range of MAP-201 is very wide, the range is 0-99.9%, and the display accuracy is very high - in addition, it is still excellent in speed and accuracy.
Technical Parameter
check the accuracy
| O2: ±0.1%(Absolute accuracy)@ O2≤10.0% ±1%(Absolute accuracy)@100%≥O2≥10.0% |
Detection mode | O2:Electrochemistry |
Detection range | O2: 0~99.9% (Resolving power:0.1%) |
response time | Adjustable,6~10 秒 |
Sensor life | Oxygen sensor for 5 years |
Minimum required sample gas | <15ml |
Gas sampling | The built-in pump inhales automatically through the testing needle |
calibration | Oxygen: 1 point calibration (air) Carbon |
user interface | 7 inch color touch screen |
Power Supply | AC220,50HZ |
ambient temperature | 0-40 Centigrade |
Ambient humidity | <95 %RH |
size | 370mmx320mmx250mm (LxWxH) |
weight | 2.0 Kg |