MAP-200 Desktop Headspace Packaging Tester
2020-05-28

1.1. Working principle

oxygen

The core component of MAP-200 is an electrochemical oxygen sensor. The principle of the sensor is similar to that of a battery. When gas passes through the sensor, the sensor can generate current corresponding to the gas concentration. The current value can be converted to the concentration value and displayed on the display screen. The sensor has excellent strength and stability, high repeatability and fast response speed. Even if the oxygen concentration is high, the sensor will not be saturated, so the application range of MAP-200 is very wide, the range is 0-99.9%, and the display accuracy is very high - in addition, the speed and accuracy are still excellent.

Carbon dioxide

CO2 sensor is a kind of independent non-dispersive infrared sensor, which is equipped with infrared light source and single wavelength filter. The sensor is a whole, without moving parts, and has a very strong structure. The range of CO2 sensor is 0-99.9%. Sensors are calibrated only once every six months, and readings are not affected by humidity or temperature.


1.2. Technical parameter

check the accuracy

 

O2

±0.1%(Absolute accuracy)@ O2≤10.0%

±1%(Absolute accuracy)@100%≥O2≥10.0%

CO2

<±2.0%@ Full range(at20℃)

Detection mode

O2:Electrochemistry

CO2:Non-dispersive Infrared Sensor(NDIR)

Detection range

O2: 0~99.9% (Resolving power:0.1%)

CO2:(0-20%,0-60%,0-99.9%)

response time

Adjustable,6~10 秒

Sensor life

Oxygen sensor for 5 years; Carbon dioxide sensor for 3 years

Minimum required sample gas

<15ml

Gas sampling

The built-in pump inhales automatically through the testing needle

calibration

Oxygen: 1 point calibration (air) Carbon dioxide: 2-point calibration (gas calibration)


user interface

7 inch color touch screen

Power Supply

AC220,50HZ

ambient temperature

0-40 Centigrade

Ambient humidity

<95 %RH

size

370mmx320mmx250mm (LxWxH)

weight

2.1 Kg

111.jpg

Prev:MAP-532 Hand-held O2/CO2 detector